Table 2. Validation factors and monitoring test for vitamin B12 in certified reference material (SRM 1849a)
| Tested value (μg/kg) | RSD (%) | Recovery (%) |
| SRM 1849a | 54.10±0.84 | 1.88 | 112.24±2.11 |
| Samples | Tested value (μg/kg) | | |
| Cereal infant formula | T-1 | T-2 | T-3 | T-4 |
| 11.93±2.08 | 11.03±0.16 | 42.18±1.57 | 16.65±1.18 |
| r2 | 0.999 | Linear regression | y=53.806x–150.44 |
| LOD | 2.71 μg/L | Range | 25–500 μg/L |
| LOQ | 8.21 μg/L | | |
All values are mean±SD of three replicates.
RSD, relative standard deviation; LOD, limit of detection; LOQ, limit of quantitation.